Show simple item record

dc.contributor.authorSherif, Mohammad A
dc.contributor.authorHerold, Joerg
dc.contributor.authorVoelker, Wolfram
dc.contributor.authorManiuc, Octavian
dc.contributor.authorErtl, Georg
dc.contributor.authorPraast, Christian
dc.contributor.authorBraun-Dullaeus, Ruediger C
dc.date.accessioned2015-09-01T16:52:55Z
dc.date.available2015-09-01T16:52:55Z
dc.date.issued2015-07-28
dc.identifier.citationBMC Cardiovascular Disorders. 2015 Jul 28;15(1):78
dc.identifier.urihttp://dx.doi.org/10.1186/s12872-015-0072-7
dc.identifier.urihttp://hdl.handle.net/10724/31674
dc.description.abstractAbstract Background Accurate preoperative assessment of the aortic annulus dimension is crucial for successful transcatheter aortic valve implantation (TAVI). In this study we validated a new method using two-dimensional transesophageal echocardiography (2D-TEE) for measurement of the aortic annulus prior to TAVI. Methods We analysed 124 patients who underwent successful TAVI using a self-expandable prosthesis, divided equally into two groups; in the study group we used the cross sectional short axis 2D-TEE for measurement of the aortic annulus and in the control group we used the long axis 2D-TEE. Results Both groups were comparable regarding the clinical parameters. On the other hand, patients in the study group had less left ventricular ejection fraction (38.9 % versus 45.6 %, p = 0.01). The aortic valve annulus was, although not statistically significant, smaller in the study group (21.58 versus 23.28 mm, p = 0.25). Post procedural quantification of the aortic regurgitation revealed that only one patient in both groups had severe aortic regurgitation (AR), in this patient the valve was implanted deep. The incidence of significant AR was higher in the control group (29.0 % versus 12.9 %, p = 0.027). Conclusions Sizing of the aortic valve annulus using cross-sectional 2D-TEE offers a safe and plausible method for patients undergoing TAVI using the self-expandable prosthesis and is significantly superior to using long axis 2D-TEE.
dc.titleFeasibility of a new method using two-dimensional transesophageal echocardiography for aortic annular sizing in patients undergoing transcatheter aortic valve implantation; a case–control study
dc.typeJournal Article
dc.date.updated2015-07-29T18:10:53Z
dc.language.rfc3066en
dc.rights.holderSherif et al.


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record